Chromatography Detectors
Product Accessories - Analytical Instruments
NPD / FTD / TID : Nitrogen Phosphorus detectors
- Temperature range settable up to 450°C. in the step of 1 °C
- Minimum Detection Level : <0.1pg N/S C/sec (azobenzene) selective for nitrogen & <0.01pg P/S C/sec (malathion) selective for phosphorus.
FID : Flame ionisation detector
- High sensitivity flame ionization detector with auto flame ignition, Temperature range settable up to 450°C. in the step of 1 °C
- Dynamic range : 107
- Minimum Detection Level : <1.2pg C/sec of dodecane
ECD : Electron captur detector
- Radioactive Source :10mCi Nickel 63
- Max Operating Temp : 350°C
- Detection Limit : 1 x 10-14 g/sec Lindane
- Linear Dynamic Range : 104 (Argon / Methane)
TCD : Thermal conductivity detector
- Dual path four Tungsten-Rhenium filaments
- Temperature range settable up to 400°C in the step of 1 °C
- Filament Current range : 0-400 mA; Temp Stability : ± 0.1°C
- Minimum Detection Level : 10 uv/ppm(vol) toluene
- Filament temperature 150 - 450 °C in steps of 75°C with filament protection circuit
FPD : Flame photometric detector